Model 2035 Czerny-Turner Monochromator
for Scanning and CCD Array Spectroscopy Experimentation
  • f/4.8 High Throughput Optics
  • High Efficiency Broadband Al+MgF2 Coatings
  • Precision Micrometer Adjustable Slits
  • Dual Turret with "Snap-In" gratings
  • Gratings rotate about their apex
  • Imaging Optics & Large focal plane
  • Rugged Construction with Multiple slit locations
Model 2035 350-mm focal length f/4.8 Monochromator

The Model 2035 Czerny-Turner spectrometer has a focal length of 350-mm and aperture ratio of f/4.8. This instrument uses a concave optic to collimate the incoming radiation. A plane grating is used to diffract the radiation that is subsequently focused by a second concave mirror. Czerny-Turner instruments require the center of the grating to be mounted directly over its center of rotation. This fixes the diffraction angle and provides the user with a fixed trajectory exit beam. This keeps accessories after the exit slit on the optical axis and delivers a flat and fixed focal plane.

The Model 2035 has a high performance optical system. The Model 2035 utilizes master concave collimating and focusing optics polished to 1/8th wave. The high quality optics provide excellent performance from the Infrared to the Deep UV. This system is limited by the atmospheric cut-off (185-nm) and is available with fittings for purge gas. Spectrally agile, the Model 2035 features first surface optics with Aluminum and Magnesium Fluoride coatings for broadband response. To favor Infrared applications other optical coatings like Gold or Silver can be provided.

Wavelengths from 200-nm to 78-um can be covered within the scanning range and with appropriate gratings. Wavelength control is via the 789A-3 digital scan control . The stepper drive (refer to Model 789A-3) is interfaced for operation via PC.

The Model 2035 optical system is optimized for two modes of operation. Instruments may be equipped with spherical optics for optimal spectral resolution. Spherical optics introduce astigmatism to the system that is not desirable in multi-channel fiber or other imaging applications. Configured for imaging the Model 2035 is equipped with an asphere to reduce astigmatism. Learn about the Model 2035 for imaging here.

PDF format Data Sheet for the high throughput Model 2035 available here

Specifications

McPherson Model 2035 350-mm focal length f/4.8 Monochromator / Spectrometer
Focal Length350-mm, Czerny Turner design Monochromator / Spectrometer
Slit LocationsAxial and lateral with optional extra entrance and exit port selection mirrors
f / No.4.8 with turret or single grating holder
Grating Size68 x 68-mm
Wavelength Accuracy+/-0.2-nm (on counter, with 1200 G/mm grating)
Wavelength Reproducibility+/- 0.005 nm (with 1200 G/mm grating)
Focal Plane25-mm, multiply dispersion by the width of your detector for range
Wavelength Rangerefer to grating of interest for range


Performance of the Model 2035 with various gratings:
Grating (G/mm) (others available) 2400 1800 1200 600 300 150 75 50
Wavelength Range from 185-nm to 650-nm 860-nm 1.3-um 2.6-um 5.2-um 10.4-um 20.8-um 78-um
First Order Littrow Blaze
240 nm
holo
250 nm
300 nm
750 nm
1.25 um
2.0 um
45 um
300 nm
300 nm
500 nm
1.0 um
2.5 nm
3.0 nm
holo
500 nm
750 nm
3.0 um
4.0 um
8.0 um
750 nm
1.0 um
4.0 um
6.0 um
10.0 um
1.0 um
1.85 um
8.0 um
12.0 um
Resolution (nm) at 313.1-nm 0.025 0.03 0.05 0.10 0.20 0.40 0.80 1.20
Dispersion (nm/mm) 1 1.5 2 4 8 16 32 48



Salient Features