Model 209 Czerny-Turner Monochromator
for High Resolution Spectroscopy
  • Exceptional Spectral Resolution
  • Patented "Snap-In" gratings
  • Imaging Optics & Large focal plane
  • Rugged Construction with Multiple slit locations
  • Extended wavelength range setting
  • Optional equipment includes Model 608 prism predisperser, oversize Echelle gratings and Double Resolution / Double Pass optics
Model 209 High Resolution Monochromator

McPherson, Inc. Research Grade Czerny-Turner Monochromators are available in many different focal lengths and with many different options. Chances are good that we have one just right for your application. Models ranging from 0.67-m to 2-meter focal lengths share a wealth of features.

Used in air the spectral range of these instruments extends from 185-nm to 78-µm, and with vacuum down to 105-nm (depends on the grating.) For intermediate ranges a purge gas can be used (Nitrogen, for example.) McPherson SNAP IN gratings allow the alignment-free use of many different gratings. Easily view a wider spectral range or obtain higher resolution! You can retrofit years later and not have to worry about alignment with the McPherson SNAP IN approach. Our dual grating turret also accepts these easy to exchange gratings.

McPherson optics are Aluminum coated with a top layer of Magnesium Fluoride. This coating provides extremely good response in a broad spectral range. If required by your application other coatings (Gold, Silver, Aluminum, etc.) can be provided.

Well equipped, these instruments feature: multiple entrance and exit ports, high precision wavelength drives, extended range operation, Echelle, and oversize grating mounting capabilities. Optional features include grating turrets, imaging optics, multi-source, and multi-detector turrets further increasing versatility. See the features sheet to learn more.

PDF format Data Sheet for the Model 209 available here

The Model 209 offers the user high spectral resolution. We also make the Model 209, and all the long focal length Czerny-Turner models, in vacuum compatible versions.

Specifications

McPherson Model 209 1.33-meter focal length f/9.4 Monochromator
Focal Length1.33-meter, Czerny Turner design Spectrometer with Patented "Snap-In" gratings
Slit LocationsAxial and lateral with optional extra entrance and exit port selection mirrors
f / No.9.4 (11.6 with smaller grating)
Grating Size120 x 140-mm (or 110 x 110-mm) - Echelle gratings up to 220-mm wide
Wavelength Accuracy+/-0.05 nm (with 1200 G/mm grating)
Wavelength Reproducibility+/- 0.005 nm (with 1200 G/mm grating)
Focal Plane50-mm maximum width, multiply dispersion by the width of your detector for range
Wavelength Rangerefer to grating of interest for range, in extended position increase top limit 20%


Performance of the Model 209 with various gratings:
Grating (G/mm) (others available)
2400
1800
1200
600
300
150
75
20
Wavelength Range from 185 nm to
650 nm
860 nm
1.3 um
2.6 um
5.2 um
10.4 um
20.8 um
78 um
First Order Littrow Blaze
240 nm
holo
250 nm
300 nm
750 nm
1.25 um
2.0 um
45 um
300 nm
300 nm
500 nm
1.0 um
2.5 nm
3.0 nm
holo
500 nm
750 nm
3.0 um
4.0 um
8.0 um
750 nm
1.0 um
4.0 um
6.0 um
10.0 um
1.0 um
1.85 um
8.0 um
12.0 um
Resolution (nm) at 313.1 nm
0.005
0.007
0.01
0.02
0.04
0.08
0.16
0.60
Dispersion (nm/mm)
0.31
0.41
0.62
1.24
2.48
4.96
9.92
37.2