Spectral Emission, Excitation and Luminescence Measurement ~ Optical Characterization
- 120 to 900nm wavelength range
- 5 sample positions
- Lifetime / Persistence measurement capability
- Fast PMT (optionally cooled) equipped with signal recovery
- Tunable monochromatic Excitation
- Tunable monochromatic Emission - optionally CCD equipped
|
|
 |
The McPherson Vacuum UV Spectrophotometer is designed for Transmission, Reflection and Emission measurements across a broad wavelength range. The system has a usable wavelength range of 115nm to 930nm - this can be expanded deeper into the Extreme UV or into the IR with addition of sources and detectors. The interchangeable sample holder(s) accept five 25mm diameter samples in the emission, transmission and reflectance positions. Samples can be indexed under vacuum. The optical, electrical and pumping systems are built into the supporting table for easy transport and set up. The computer sits on a separate table. All cables and lines are quick disconnect type and are labeled for easy set up.
Excitation and Emission The excitation source is a CW
Deuterium lamp with Magnesium Fluoride (MgF2) window. The lamp window seals the instrument vacuum vessel and all emitted wavelengths from about 120nm to 380nm can be used for excitation. The discrete excitation wavelength is selected by an optimized Vacuum UV McPherson
Model 234/302 high throughput 200mm scanning monochromator. Monochromatic excitation energy is
collected and focused to the sample position. Alternate means of excitation are possible by use of accessory ports in the sample chamber, Excimer laser or x-ray sources for example.
Emission spectra is detected with a McPherson
Model 218 high resolution 300mm scanning monochromator. It can scan from 105nm to far IR spectral region. If emission spectra are always >200nm, we can substitute the
Model 2035 high resolution scanning spectrometer. The sample emission is collected and focused to the entrance slit. Many types of solar blind and/or cooled photomultiplier detectors can be fitted to the exit as can CCD or intensified CCD detectors.
When measuring reflection or transmission the excitation and emission monochromators can be scanned synchronously or one can be positioned at zero order. Scanning both instruments with an offset to reject fluorescence is also possible.
Specifications
Vacuum UV Optical Characterization System
| Excitation Wavelength(s) | < 1nm FWHM discrete bands selectable from 120 to 380nm Vacuum monochromator Model 234/302 Filters and Excimer sources - optional |
| Emission Wavelength(s) | < 1nm FWHM discrete bands measured from 120 to 900nm PMT used for lifetime and scanning CCD for Color and survey work - optional |
| Sample Size | 5 ambient temperature 25mm diameter samples vertical mounting |
| Sample Environment | Ambient / Room temperature Cryogenically cooled (single sample) - optional Heated (to 400deg C) (single sample) - optional |
Mounting Samples Optics in the sample chamber focus the energy from the excitation monochromator. The focused energy is transmitted through or reflected off the samples within the chamber and refocused into the emission monochromator. The sample chamber has two sample holder positions, a transmission position and a reflectance/emission position. The transmission position places the samples normal to the excitation beam, a first surface mirror must be placed in the reflectance/emission position during transmission measurements. The reflectance/emission sample holder position places the samples in the beam focus and allows the samples to be rotated with respect to the excitation beam angle of incidence. This angle adjustment can be used to exclude the specular component when measuring fluorescent or diffuse samples. The angle adjustment can also be used to optimize the signal when measuring glossy samples such as mirrors or wafers.
Emission mode signal collection optics can be tuned to optimize light collection and focus at the emission monochromators entrance slit while under vacuum.

Sample mounts heated to 400° C or cooled to LN2 temperatures can be quoted on request.
Optional auxiliary source and detector ports are available for use and may be configured using movable mirrors. Access to the mirrors is gained through the access ports.
Time Resolved Spectra / Lifetime
| Application Example: Set the vacuum ultraviolet spectrometer to desired Excitation wavelength. Set the Emission monochromator to 525 nm. Select a (green) emitting sample. Enter 525nm in the GO TO parameter field of the software and then press GO TO. The monochromator will proceed to 525nm and data acquisition will commence. The McPherson systems uses a SR430 Multi-Channel Scalar for measurement of lifetime / decay. The function generator DS335 is set to produce a trigger pulse. The trigger pulse from the DS335 function generator closes a shutter and starts acquisition by the SR430 Multi-Channel Scalar. |
|
 |
Spectral Emission, Excitation Measurement
 | |
Application Example: Set the vacuum ultraviolet spectrometer to desired Excitation wavelength. Set the Emission monochromator to SCAN from 350 to 600nm. Select a (green) emitting sample. The McPherson systems uses a SR430 Multi-Channel Scalar for measurements during Emission wavelength scans. This mode will take the mean data value and record one data point per wavelength increment. Enter the desired wavelength BEGIN, END and INCREMENT values in software and then SCAN. The monochromator scan will commence and data acquisition will begin. The screen will update until all wavelength increments have been recorded. |
Reference Detector A small diverter mirror directs a small amount of energy from the excitation monochromator to a scintillated photomultiplier. This detector monitors the intensity of the excitation source. Stability information may be fed back into the data to eliminate source fluctuation from the data. Calibrated PMT detectors can be used in this position. The mirror can be switched in or out of the beam via a lever in the top cover of the sample chamber.
Vacuum Pumping System The system is safety interlocked and designed to shut down all pumps and vent the system in the event of a problem. The main components in the system are two 250 liter/sec turbo pumps backed by two 3.2 CFM mechanical pumps. The vacuum is monitored by inverted magnetron and thermocouple gauges. The vacuum controller has several safety features to protect the system. The main danger is contamination of the optical system. The two common causes of contamination are either running the mechanical pump without the turbo or running only one of two turbos. If the controller detects either of these conditions, all pumps are shut down and the system is vented. The controller also shuts down and vents in the event of power failure. Backfill with dry nitrogen when venting.
Software Controls Complete, easy to use software for data collection from sample and reference detectors is combined with our Spectrometer Control package. In addition to controlling the scan wavelength range, go to position and wavelengths for the spectrometers all features as used with detection equipment are also controlled. Easy to use, point and click, software for control of SRS400 photon counter, 430 5nsec multi-scaler averager, 830 lock-in amplifier, EG&G PAR 5209 lock-in amplifier, Keithley 4853 picoammeter and ADC16 has been developed and will work with any of our dual spectrometer systems. Other devices may be controllable - please inquire.